OSA Warnings
Table 5-6. OSA Warnings (8 of 12)
Error Number
Error Description [description/explanation/examples]
6700
desc = "Math expression input parameter undefined."
help = "A math expression could not be evaluated because one or more input
arguments are undefined. Please check the spelling of all input arguments."
6701
desc = "Math expression input parameter has error."
help = "A math expression could not be evaluated because an error is present on
one or more of the input arguments. The error must be cleared before the
expression can be evaluated."
6702
desc = "Math expression parameter has zero length."
help = "A math expression could not be evaluated because one or more input
arguments has a zero length. This is often caused when the subset of a trace
contains no points such as when advanced line marker functions are turned on
and the line markers are too close together. It can also be caused when peak/pit
searches find no peaks or pits."
6720
desc = "Math expression input parameter has error."
help = "A math expression could not be evaluated because of improper input
arguments. This could be due to one or more input arguments having the wrong
type or size or because one or more inputs are not defined."
6721
desc = "Internal error in marker search"
help => "A marker search function has failed due to an internal software
problem. Please try the operation again. If the error persists, contact the
Agilent support center nearest your location."
6722
desc = "Math expression input cannot be boolean."
help = "A math expression could not be evaluated because one or more of the
input arguments is boolean. This function does not allow boolean input
arguments."
6723
desc = "Trace X axis values do not match."
help = "A math expression could not be evaluated because the input arguments
have differing X axis values. This function requires that all input arguments have
identical X axis values. All trace inputs must be taken with identical start/stop
settings."
6724
desc = "Error in source trace"
help => "The source trace has an error. The error may possibly be that the source
trace contains no points. Please try the operation again. If the error persists,
contact the Agilent support center nearest your location."