Calibrating Wavelength Measurements

Calibrating Wavelength Measurements

Environmental variations such as air pressure, temperature, and humidity can
affect the index of refraction of air in the monochromator of the optical spec-
trum analyzer (OSA). This section discusses calibration methods that you can
use to improve the wavelength accuracy in the Agilent 8614X OSA's.
"Calibration" on page 3-17
Many aspects of remotely programming the optical spectrum analyzers are discussed
in Product Note 86140-2R, Wavelength Calibration for the 86140X Series Optical
Spectrum Analyzers (Literature part number 5980-0043E).
Wavelength calibration routines improve wavelength accuracy by determining
errors and correcting them with offsets, using linear interpolation when neces-
sary. For maximum wavelength accuracy, calibration points spaced a maxi-
mum of 10 nm apart are recommended.
You can perform a wavelength calibration by using one of the following meth-
ods:
These calibration routines should only be performed after the instrument's
temperature has been stabilized by a minimum of 1 hour of continuous opera-
tion.