Function Reference
Sweep Points
Span
Stop
=
Center
+
------------ -
2
Use the knob, step keys, or numeric keys to enter the desired value.
If the instrument is in zero span, this command sets the center wavelength to
the value specified.
The default for stop wavelength is 1700 nm.
Wavelength > Stop WL
Span, Start WL, Center WL
SENSe:WAVelength:STOP
Sweep Points
Specifies the number of data points taken for a sweep. The more data points
the better the trace resolution, but the longer the sweep time. You can select
from 3 to 10001 points. Enter the number of data points using the step keys,
numeric key pad, or knob. The default is 1001 points.
Traces > Trace Setup > Sweep Points
SENSe:SWEep:POINts