Normal/Delta Marker Interpolation (Marker Setup Panel)
Normal/Delta Marker Interpolation (Marker Setup
Panel)
Turns the normal/delta marker interpolation on or off. When on, the normal/
delta markers will be placed at the exact wavelength or frequency value
selected if it is within the trace range. The marker will linearly interpolate
between two trace data points. When off, the marker will be placed on the
nearest trace data point.
This setting enables interpolation for all four markers when in normal or delta
mode.
Markers > Marker Setup > Normal/Delta Marker Interpolation
Active Marker, Delta Marker
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:INTerpolation
Normal Marker Units (Marker Setup Panel)
Sets the marker X-axis readout for frequency or wavelength when the instru-
ment is in a non-zero span. The X-axis default units is nm (nanometers). This
setting controls only the normal marker X-axis readout and the delta refer-
ence readout. This setting controls all four normal markers. The bandwidth
and delta offset markers have their own setting.
Markers > Marker Setup > Normal Marker Units
Active Marker
CALCulate[1|2|3|4|5|6]:MARKer[1|2|3|4]:X:READout