Specifications and Regulatory Information
Specifications
Amplitude
Flatness
k
1290-1330 nm
1525-1570 nm
1525-1610 nm
1250-1610 nm
l
Polarization Dependence
1310 nm
1530 nm, 1565 nm
1600 nm
1250 to 1650 nm
a. The 86144B and 86146B specifications are for the 50
b. Sensitivity is defined as signal value >6 X RMS noise value
°
c. Temperature range 0 to 30
d. Second Order
°
e. Temperature range 20 to 30
f. Resolution bandwidth setting <channel spacing
g. Characteristic
≥
h. For resolution
0.1 nm
i. Excluding amplitude errors at low power levels due to noise
j. For signals within 8 dB of top of screen
k. With applied input fiber 9/125
l. Absorption of light by atmospheric moisture affects flatness at 1350-1420 nm
≥
m. For resolution
0.2 nm
n. Room temperature
7-8
Agilent 86140B, 86143B,
86144B
-----
±
0.2 dB
-----
-----
k,
,
m
n
±
0.25 dB
±
0.2 dB
±
0.25 dB
±
0.3 dB
µ
m internal path only.
C
C
µ
m
Agilent 8614xB
a
Agilent 86142B, 86145B,
86146B
±
0.2 dB
-----
±
0.2 dB
±
0.7 dB
±
0.12 dB
±
0.05 dB
±
0.08 dB
±
0.25 dB
Agilent 86141B,
86140B W/Opt 025,
86143B w/Opt 025
-----
±
0.2 dB
-----
-----
----
----
----
±
0.5 dB