Agilent Technologies 86142B User Manual page 211

Optical spectrum analyzer
Table of Contents

Advertisement

Status Listings
SCPI-Defined Errors
Table 5-2. Execution Errors (2 of 3)
Error Number
Error Description [description/explanation/examples]
–222
desc = "Data out of range"
help = "A numeric value was entered which is outside the legal range of values
for the parameter. The name of the parameter is listed at the end of the error
message."
[Indicates that a legal program data element was parsed but could not be
executed because the interpreted value was outside the legal range as defined
by the device (see IEEE 488.2, 11.5.1.1.5.)]
–222
desc = "Span out of range"
help = "A numeric value was entered which is outside the legal range of values
for the span setting. This occurs if the value is too large less than zero or
between zero and 0.2nm. Zero span may not be entered by using the DOWN
arrow key from 0.2nm; a numeric value of zero must be entered instead."
[Indicates that a legal program data element was parsed but could not be
executed because the interpreted value was outside the legal range as defined
by the device (see IEEE 488.2, 11.5.1.1.5.)]
–223
desc = "Too much data"
help = ""
[Indicates that a legal program data element of block, expression, or string type
was received that contained more data than the device could handle due to
memory or related device-specific requirements.]
–224
desc = "Illegal parameter value"
help = ""
[Used where exact value, from a list of possibles, was expected.]
–257
desc = "File name error"
help = ""
[Indicates that a legal program command or query could not be executed
because the file name on the device media was in error; for example, an attempt
was made to copy to a duplicate file name. The definition of what constitutes a
file name error is device-specific.]
–272
desc = "Macro Exec Error"
help = ""
[Indicates that a syntactically legal macro program data sequence could not be
executed due to some error in the macro definition (see IEEE 488.2, 10.7.6.3.)]