Table 5-9. OSA Errors (2 of 2)
Error Number
Error Description [description/explanation/examples]
20005
desc = "Trans-Impedance calibration data is invalid"
help = "Factory calibration data for the trans-impedance amplifier is invalid.
Please record the hexadecimal number listed with the error and cycle power. If
the error persists, contact the nearest Agilent Technologies Instrument support
center for assistance. In the U.S., call (800) 403-0801. See the Agilent 86140B
Series Users Guide for a listing of the Agilent sales and service offices."
20006
desc = "Monochromator flatness calibration data is invalid"
help = "Factory flatness data for the monochromator is invalid. Please record the
hexadecimal number listed with the error and cycle power. If the error persists,
contact the nearest Agilent Technologies Instrument support center for
assistance. In the U.S., call (800) 403-0801. See the Agilent 86140B Series Users
Guide for a listing of the Agilent sales and service offices."
20007
desc = "Sweep Timeout"
help = "A sweep was started but did not finish in the expected amount of time.
The trace data acquired during this sweep may not be valid. Try taking another
sweep. If the error persists, contact the nearest Agilent Technologies Instrument
support center for assistance. In the U.S., call (800) 403-0801. See the
Agilent 86140B Series Users Guide for a listing of the Agilent sales and service
offices."
21999
desc = "The Error list has overflowed"
help = "The Error list has overflowed. The last entries received have been
deleted."
Status Listings
OSA Errors
5-37