2.5 Grounding Requirements
2.5.2 System Low-Impedance Grounding
Within the vacuum system, the low-impedance ground is provided by 3"-wide and 1/2"-wide
copper straps. As Figure 2-3 shows, these straps must connect:
the grounding stud labeled RF GND on the power module's rear panel (see Figure 2-4) to a
grounding point on the frame of the operator station
the operator station's grounding point to the vacuum cubicle's central grounding point
the grounding stud labeled RF GND on the FPS front panel (see Figure 2-) to the vacuum
cubicle's main grounding point
the source tray to the vacuum cubicle's main grounding point.
Figure 2-3 shows the low-impedance ground connections for a single-gun system. Low-
impedance ground connections are identical on two-gun units, except that an additional
length of ½"-wide copper strap must connect the system's central grounding point to FPS #2.
For an additional diagram relating to low-impedance grounding, see either document 0101-
8300-0, the Quick Start Guide for single-gun CV-12SLX units, or document 0101-8300-1, the
Quick Start Guide for dual-gun units.

Figure 2-3 System Low Impedance Grounding Requirements

2.5.3 Power Module Grounding
Make connections to the power supply's grounding lugs as shown in the Quick Start Guides for
the CV-12SLX, following the procedure described below.
CV-12SLX User Manual
2-6
Section 2: Installation