2.6 Cable Installation

Figure 2-7 Input Power Connections to 208-Volt HVPS Units

On 400-V units, input power wiring must meet the following specifications:
30 A, 2-Phase wye, 5-conductors (including neutral), each #10 AWG stranded UL1015 wire
(Also consult local electrical codes.)
Wire the conductors in this cable into the input power terminal connector supplied with the
HVPS (see Figure 2-), so that the three hot wires will connect to rear panel receptacles L1,
L2, and L3, the neutral wire will connect to rear panel receptacle N, and the ground wire will
connect to the receptacle marked as a ground. Then plug the terminal connector into the rear
panel INPUT POWER receptacle.

Figure 2-8 Input Power Connections to 400-Volt HVPS Units

Connecting the FPS Input Power Cable(s) (PN 6622-0100-20)
An FPS power cable (PN 6622-0100-20) is provided for each FPS. In one-gun installations,
plug one end of this cable into HVPS rear panel connector J1 (see Figure 2-4). In two-gun
installations, plug the power cable for FPS #1 into J1 and the power cable for FPS #2 into J2.
Plug the other end of the power cable into the power receptacle on the FPS front panel (see
Figure 2-5).
Make sure that the ON/OFF switch on the FPS front panel (see Figure 2-)
remains in the OFF position throughout the installation procedure.
CV-12SLX User Manual
CAUTION
2-10
Section 2: Installation
0101-8292-0, Rev. C