3.5 Control of Power Supply from a TemEBeam Controller (EBC)
Figure 3-3 EBC's Operations>E-Beam Screen when Beam Is Off
3.5.2 Using the E-BEAM ON/OFF Button to Switch the Beam On/Off
As noted above, when the beam is off you can switch it on by simply touching the E-BEAM
button, assuming that you have already entered a nonzero percent-power setpoint. This applies
when both HV and gun are off, when the HV alone is on, and when the gun alone is on.
Touching the E-BEAM button when the beam is on switches the HV and the gun off. Figure 3-4
shows the Operations>E-Beam screen when the beam is on.
Figure 3-4 Operations>E-Beam Screen, Beam On at 0.0 mA
3.5.3 Switching on the HV and Gun Independently of Each Other
Figure 3-5 shows the Operations E-Beam screen when the gun is on but the HV is off. To switch
the HV on/off independently of the gun, simply touch the HV ON button,. Likewise, to switch
the gun on/off independently of the HV, touch the GUN ON button. You can also switch on the
beam by touching the HV ON and GUN ON buttons in either order.
CV-12SLX User Manual
3-4
Section 3: Power Supply Operation
0101-8292-0, Rev. C