Section 2: Installation
2.6.5 Making I/O Connections to the Vacuum System
The installer must make connections (1) between the vacuum system and the HVPS I/O cable
(PN 0620-6682-0) and (2) between the vacuum system and the FPS I/O cable (PN 0620-6672-
2). Control inputs and status outputs include both digital and analog signals. Digital inputs
require a simple contact closure and carry 24 V dc at approximately 20 mA. All digital outputs
are +24 V dc sourcing a maximum of 50 mA. For detailed information about the I/Os
exchanged via the HVPS I/O cable, see Figure 2-16 and Table 2-1. For details about I/Os
exchanged via the FPS I/O cable, see Figure 2-17 and Table 2-2.
When making I/O connections to the HVPS I/O cable, use either the optional 25-pin interface
module available from Temescal (PN 6149-2293-637) or a comparable interface module.
When making I/O connections to the FPS I/O cable, use either the optional 15-pin interface
module available from Temescal (PN 6149-2293-624) or a comparable interface module.
0101-8292-0, Rev. C
3
Plug the male end of the same HVPS I/O cable (PN 0620-6682-0)
into HVPS rear panel connector J4 (see Figure 2-4).
4
Plug the male end of the FPS I/O cable (PN 0620-6672-2) for Gun
#1 into connector J101 on the front panel of FPS #1 (see Figure 2-
15). If this is a single-gun installation, skip Steps 5-7 and proceed to
Step 8 of this procedure. If you are performing a dual-gun
installation, you must also complete Steps 5-8 below.
5
Plug the male end of the HVPS-FPS cable (PN 0620-6672-0) for Gun
#2 into connector J6 on the HVPS rear panel (see Figure 2-5).
6
Plug the female end of the same HVPS-FPS cable into connector
J102 on the front panel of FPS #2 (see Figure 2-15).
7
Plug the male connector on one end of the FPS I/O cable (PN 0620-
6672-2) for Gun #2 into front panel connector J101 on FPS #2 (see
Figure 2-15).
8
Make the I/O connections to the control system, as described in
section 2.6.5.
Before performing this portion of the installation procedure, perform the
mandatory baseline test described in Section III of either document 0101-
8300-0, the Quick Start Guide for single-gun CV-12SLX units, or document
0101-8300-1, the Quick Start Guide for dual-gun CV-12SLX units.
2-15
NOTE
2.6 Cable Installation
CV-12SLX User Manual