5.4 Troubleshooting Procedures
Before performing the following procedure, perform the procedure described in
section 5.4.1.
Condition indicated: Inverter current > 150 A
Symptoms
TCS displays PSU1_FAIL_ALM (= Power Supply Control Is On failure alarm)
HVPS Front Panel: INVERTER OVERLOAD and POWER SUPPLY FAULT LEDs are lit
EBC: Fault LED lit in High Voltage section of Diagnostics screen
Possible Cause
1. Incorrect or fluctuating line voltage
2. Faulty or loose connections.
If preceding steps have failed to correct the fault, perform the following procedure:
1.
Switch off the power supply.
2.
Disconnect HV output cable from HVPS J1.
3.
Switch on the power supply and the HV and leave them switched on for no more than 9 seconds, observing whether the
symptom persists during that interval.
DO NOT leave the high voltage on for more than 9 seconds with the HV cable disconnected. Doing so will
result in the emitter's filament burning out.
4.
If the symptom is not in evidence during Step 3, troubleshoot the emitter assembly, following the instructions in the e-gun
manual. If the symptom persists during Step 3, reconnect the HV output cable to HVPS J1, switch the power supply back
on, and proceed with the troubleshooting steps described below.
Possible Cause
3. Rail voltage incorrect
4. Fault condition(s) whose investigation
requires removal of HVPS top cover
CV-12SLX User Manual
DANGER: HIGH VOLTAGE

Table 5-17 HVPS INVERTER OVERLOAD LED lights red

1. Confirm input line voltage
2. Make sure that all HV and FPS connections are correctly and securely made.
3. See Table 5-12.
4. Contact Temescal Field Service.
Procedure
CAUTION
Procedure
5-26
Section 5: Troubleshooting
0101-8292-0, Rev. C