Section 2: Installation
2.5.2 System Low-Impedance Grounding
Within the vacuum system, the low-impedance ground is provided by 3"-wide and 1/2"-wide
copper straps. As Figure 2-3 shows, these straps must connect:
the grounding stud labeled RF GND on the power module's rear panel (see Figure 2-4) to a
grounding point on the frame of the operator station
the operator station's grounding point to the vacuum cubicle's central grounding point
the grounding stud labeled RF GND on the FPS front panel (see Figure 2-5) to the vacuum
cubicle's main grounding point
the source tray to the vacuum cubicle's main grounding point.

Figure 2-3 System Low Impedance Grounding Requirements

2.5.3 Power Module Grounding
Make connections to the power supply's grounding lugs as shown in the CV-6SLX Quick Start
Guide (PN 0101-8241-1), following the procedure described below.
0101-8242-0, Rev. C
Step
Action
1
A roll of 1/2"-wide copper strap (PN 5621-0032-0) is supplied with
the unit. Cut off a length of this strap that will easily extend from
the grounding lugs on the power module's rear panel to the
operator station's frame.
2-5
2.5 Grounding Requirements
CV-6SLX User Manual