2.5 Grounding Requirements

2.5
Grounding Requirements
2.5.1 Facility Low-Impedance Grounding Requirements
Safe, dependable operation of the power supply cannot be ensured unless a good earth
ground is provided for the system and the power supply. This ground must provide a low-
impedance path for radio frequency (RF) as well as direct current (dc) electricity, and it must
not be connected to that of any other system or equipment. Figure 2-2 shows two different
methods of providing the required low-impedance ground on the facility side of the
installation.
CV-12SLX User Manual

Figure 2-1 Mounting Hole Pattern for FPS Module(s)

2-4
Section 2: Installation
0101-8292-0, Rev. C