Section 5: Troubleshooting

Figure 5-3 Sub-Threshold Arcing: Arc/Out Of Regulation and Output Arcing LEDs

Arc/Out of Regulation LED
LED is normally off (i.e., clear)
Flashes yellow when subthreshold
arcing occurs and when the HV is
out of regulation
HV Out of Regulation LED
LED is normally green
Turns amber for 2 seconds each
time the Arc/Out of Regulation
LED flashes amber

Figure 5-4 Filament Load Fault and Filament Power LEDs

Filament Load Fault LED
LED is normally green when FPS
is powered up
Turns red to indicate an 'open'
on the filament circuit (i.e.,
filament broken or burned out;
resistance across emitter is 2.5
Ohms or greater)
LED is green when FPS is
powered up, red when FPS is
not powered up
The HV can remain on, even if the gun is switched off, either due to a fault or user
action.
5.2.3 LEDs That Indicate Latching Faults
Figures 5-6 through 5-10 point out and describe the front panel LEDs that indicate latching
0101-8292-0, Rev. C
Figures 5-4 and 5-6 through 5-10 do not show HVPS LEDs as they would
appear when the applicable faults have occurred. Instead, the LEDs
described in these illustrations are shown as red merely to highlight
them.
Filament Power LED
NOTE
NOTE
5-3
5.2 HVPS Front Panel LEDs