Read Commands (Data Requested From The Thermostat) - Lauda PROLINE Edition X P 5 C Operating Instructions Manual

Hide thumbs Also See for PROLINE Edition X P 5 C:
Table of Contents

Advertisement

Proline Edition X heating thermostats
START
STOP
RMP_SELECT_X
RMP_START
RMP_PAUSE
RMP_CONT
RMP_STOP
RMP_RESET
RMP_OUT_00_XXX.XX_XXXXX_
XXX.XX_X
RMP_OUT_02_XXX
Permitted data formats:
-XXX.XX
-XX.XX
-X.XX
-.XX

8.3.6 Read commands (Data requested from the thermostat)

Command
IN_PV_00
IN_PV_01
IN_PV_03
IN_PV_04
IN_PV_05
IN_PV_10
IN_PV_13
IN_SP_00
IN_SP_01
IN_SP_02
IN_SP_03
IN_SP_04
IN_SP_05
IN_PAR_00
IN_PAR_01
IN_PAR_02
IN_PAR_03
IN_PAR_04
IN_PAR_05
YACE0099 / 10/9/2018
 For "_" use also " " (blank character).
 Response from thermostat "OK" or in case of error "ERR_X" (RS-485 interface
e.g. "A015_OK" or in case of error "A015_ERR_X".).
-XXX.X
-XXX.
-XX.X
-XX.
-X.X
-X.
-.X
.XX
Read bath temperature (outflow temperature)
Indication of the controlled temperature (internal/external Pt/external
Analogue/external Serial)
Read external temperature TE (Pt100)
Read external temperature TE (Analogue input)
Read current bath level
Read bath temperature (outflow temperature) in 0.001 °C
Read external temperature TE (Pt100) in 0.001 °C
Read temperature setpoint
Read current pump power level
Read cooling operation mode (0 = OFF / 1 = ON / 2 = AUTOMATIC)
Read current overtemperature switch-off point
Read current outflow temperature limit TiH
Read current outflow temperature limit TiL
Read control parameter Xp
Read control parameter Tn (181 = OFF)
Read control parameterTv
Read control parameter Td
Read control parameter KpE
Read control parameter TnE (980 = OFF)
Interface modules
Switches the device on (from Standby). See safety information
( 7.7.3).
Switches the device into Standby (pump, heater off)
Selection of the programme (1 — 5) to which the further instruc-
tions apply. When the unit is switched on, programme 5 is se-
lected automatically.
Start the programmer
Hold (pause) the programmer
Restart the programmer after pause
Terminate the programmer
Delete the programmer (all Segments)
Set a programme segment (temperature and time, tolerance and
pump level). A segment is added and appropriate values are
applied to it.
Number of times the programme runs: 0 = unlimited / 1 — 250.
-XXX
XXX.XX
-XX
XX.XX
-X
X.XX
.X
XXX.X
XXX.
XX.X
XX.
X.X
X.
Explanation
XXX
XX
X
93