X-Ray Generator And X-Ray Tube Assembly - AGFA DR 100e User Manual

Digital radiography
Hide thumbs Also See for DR 100e:
Table of Contents

Advertisement

X-ray generator and X-ray tube assembly

Table 13: X-ray generator and X-ray tube assembly
Model name
Rotating anode monobloc MHF2030
High voltage generator
Thermo-tinned structure
Max. power (100 kV - 320
mA)
Max. voltage to the tube
Max. current to the tube
Ripple to the max. power
Rise time to the max.
power
Performance
According to the load curves of the x-ray tube
Housing
Min. inherent filtration
@75 kV
Dimensions (W x D x H)
Weight
Thermal features
Thermal capacity
Thermal safety
Thermal switch
Lung
Continuous thermal dissi-
pation
Max. housing tempera-
ture
Heating curves (50 W, 100 W, 200 W, 300 W) and cooling curve (***)
DR 100e (Digital Radiography) | Product Information | 113
32 kW
125 kV
450 mA
< 1%
< 2 ms
1.1 mm Al
320 mm x 140 mm x
241 mm
19.4 kg
500 kJ
60 °C ± 5° C
normally closed
3
0.16 dm
55 W
60 °C
324.5 mm x 145 mm x
244.5 mm
18.7 kg
600 kJ
3
0.2 dm
60 W
0365B EN 20181015 0853

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6012/0016012/0026012/0036012/004

Table of Contents