The key for constant current operation is the maximum allowable voltage for the MX
can supply the UUT above the set voltage. This value is the OVOLT. The OVOLT value
must be above the set voltage for the power source. If this value is left at the set voltage
or lower, the MX will not operate at constant current in regenerative mode. The UVOLT
is used to shut the UUT if the MX could not maintain the current at set CURR value in
the regenerative screens after a time specified by the DELAY. If the UVOLT will not
control the shutdown of the UUT, then this value should be set the same as the
OVOLT. This will eliminate excessive current before the shut down.
9.9.8
REGENERATE Control Screen
The –SNK option mode of operation is controlled by parameters set in the REGENERATE
CONTROL SCREENS. Access to this screen is from the APPLICATIONS screen by selecting
the REGEN entry and pressing the ENTER key.
MX15

Figure 9-24: REGENERATE CONTROL screen

167