Ametek California Instruments MX15 Series User Manual page 170

Ac and dc power source
Table of Contents

Advertisement

Number
Message String
-203
"Command protected"
-210
"Trigger error"
-211
"Trigger ignored"
-213
"Init ignored"
-220
"Parameter error"
-221
"Setting conflict"
-222
"Data out of range"
-223
"Too much data"
-224
"Illegal parameter value"
-226
"Lists not same length"
-241
"Hardware missing"
-254
"M edia full"
-255
"Directory full"
-256
"File name not found"
-257
"File name error"
-283
"Illegal variable name"
-300
"Device specific error"
-311
"M emory error"
-314
"Save/recall memory lost"
-315
"Configuration memory
lost"
Cause
Command is locked out
Problem with trigger system.
Trigger request has been
ignored.
Initialization request has been
ignored
Parameter not allowed.
Requested setting conflicts with
other setting in effect.
Parameter data outside of
allowable range.
M ore data received than
expected
Parameter value is not
supported
One or more transient lists
programmed has different
length.
N/A
No storage space left to save
settings or data.
Too many waveform directory
entries
W aveform requested not in
directory
Incorrect filename
V ariable name illegal.
Hardware related error
W aveform memory checksum
error.
User setup register contents
lost
Hardware configuration settings
lost.
Remedy
Some commands are supported by
the unit but are locked out for
protection of settings and are not
user accessible.
Unit could not generate trigger for
transient execution or measurement.
Trigger setup incorrect or unit was
not armed when trigger was received.
Check transient system or
measurement trigger system settings.
Unit was told to go to armed state but
was unable to do so. Could be caused
by incorrect transient system or
measurement acquisition setup.
Incorrect parameter or parameter
value. Check programming manual
for allowable parameters
Check other settings. e.g. trying to
program a DC offset while in A C mode
Check programming manual for
allowable parameter values
Check programming manual for
number of parameters or data block
size
Check programming manual for
correct parameters
A ll lists must be of same length or
transient cannot be compiled and
executed.
N/A
Delete other settings or data to make
room.
Delete one or more waveforms from
waveform memory to make room.
Check waveform directory for
waveform names present.
Too many or non A SCII characters
used in waveform file definition.
Use A SCII characters only
Check hardware for proper operation.
M ay be the result of incomplete user-
defined waveform download. Check
interface and try downloading
waveform again. Successful download
may clear this error condition.
A lternatively, use TRA C:DEL:A LL
command to clear waveform memory.
Store setup in same register again.
Contact service department at
service@programmablepower.com
obtain instructions on restoring
configuration data.
to
170

Advertisement

Table of Contents
loading

Table of Contents