4.2.9
CONFIGURATION Menu
The configuration menu may be used to configure various aspects of the instrument such as the
serial port (including USB and LAN), IEEE-488/GPIB address and the power-on settings of the
supply.
ENTRY
A DDRESS
BA UD RA TE
PON STA TE
LA Network
MX15 Series
DESCRIPTION
0 - 31
Sets the selected IEEE / GPIB bus address for the optional
IEEE/GPIB interface.
Factory default is address 1. The shuttle knob or the keypad
can be used to set a value from 0 through 31. Do not use
address 0 as this address is typically reserved for the GPIB
controller.
9600
Sets the baud rate for the RS232 communications port.
19200
Factory default is 38400 baud. A vailable settings are 9600
38400
through 115200 baud.
57600
The same setting is used for USB and LA N modes. For use with
115200
either USB or LA N, the baud rate in this screen must be set to
230400
460800.
460800
Note: The shuttle knob can be used to scroll through these
selections.
REG0
Determined power on state. This setting selects either non-
RST
volatile REG0 to be recalled automatically at power-on or
factory default (RST). Factory default is RST, which recalls the
factory settings. See below.
Factory default settings are:
Output relay
V oltage Range
M ode
A C
V oltage
0 V rms
Frequency 60 Hz
Note that to use REG0 for power-on default, the contents of
the register must be programmed first. See section 4.2.8. If an
empty register is selected, the power source will revert back to
RST (factory setting).
LA N
If the –LA N option is installed; pressing Enter while the cursor
is on the LA Network entry provides access to the LA N interface
setting screens listed below.
IP A ddress
Displays the IP address setting. This value can be changed by
pressing the SET key and entering a new value from the
keypad. Use the numeric data pad to enter each field. To move
between the four fields, use the decimal point key on the
keypad.
To set a fixed IP address, press SET and enter the desired IP
Open
Low