2.6.2
-XV Option Specifications
Consult factory for -XV option specifications.
2.6.3
-HF Option Specifications
The -HF option extends the maximum available output frequency from 819 Hz to 900 Hz. Some
restrictions are in effect at this increased output frequency level.
All other specifications of the MX15 system remain unchanged if this option is installed except as
noted in the table below.
Range:
Resolution:
A ccuracy:
A ccuracy:
High V oltage Range
Low V oltage Range
-HV V oltage Range
MX15 Series
-HF Option:
Frequency
-HF option:
16 Hz - 900 Hz
0.01 Hz
< from 16.00 to 81.92 Hz
0.1 Hz
> from 82.0 to 819.2 Hz
1 Hz
> from 820 to 900 Hz
± 0.01 %
Phase
16 - 100 Hz:
< 1.5°
100 - 500 Hz:
< 2°
500 – 819 Hz:
< 4°
819 – 900 Hz:
< 5°
Voltage
M aximum voltage at 900 Hz is 290 V rms
M aximum frequency at 300 V rms is 875 Hz
See Figure 2-3 -HF Option V oltage / Frequency Rating 300V range
M aximum voltage at 900 Hz is 145 V rms
M aximum frequency at 150 V rms is 875 Hz
See Figure 2-3 -HF Option V oltage / Frequency Rating 300V range
M aximum voltage at 900 Hz is 386 V rms
M aximum frequency at 400 V rms is 875 Hz .
Note: If the voltage or frequency settings shown here are exceeded for any length of
time (> 1 sec), the M X may shut down generating an over temperature fault to protect
itself.