Ac and dc power source
Table of Contents

Advertisement

2.1.3
AC Measurements
Measurement specifications apply to MX15-1 / MX15-1Pi. See notes for other models and
configurations.
Parameter
Frequency
RM S V oltage
RM S Current
Peak Current
V A Power
Real Power
Power Factor
(>0.2kV A )
Note: A ccuracy specifications are valid above 100 counts. For current and power measurements, specifications apply from
2% to 100% of measurement range.
Note: Power factor accuracy applies for PF > 0.5 and V A > 50 % of max.
MX15 Series
Range
16.00 - 820.0 Hz
0.01% + 0.01 Hz
0 - 300 V olts
0.05V + 0.02%, <100 Hz
0. 1V + 0.02%, 100-820 Hz
0 - 150 A mps
0.15A + 0.02%, <100 Hz
0. 3A + 0.02%, 100-820 Hz
0 - 400 A mps
0.15A + 0.02%, <100 Hz
0. 3A + 0.02%, 100-820 Hz
0 - 15 K V A
30 V A + 0.1%, <100 Hz
60 V A + 0.1%, 100-820 Hz
0 - 15 K W
30 W + 0.1%, <100 Hz
60 W + 0.1%, 100-820 Hz
0.00 - 1.00
0.01, <100 Hz
0.02, 100-820 Hz
Accuracy (±)
Resolution
0.01 to 81.91 Hz
0.1 to 500 Hz
0.01 V olt
0.01 A mp
0.01 A mp
10 V A
10 W
0.01
15

Advertisement

Table of Contents
loading

Table of Contents