Options; Introduction; Option -Hv: Additional Ac Voltage Range - Ametek California Instruments MX15 Series User Manual

Ac and dc power source
Table of Contents

Advertisement

9 Options

9.1 Introduction

There are a number of options available for the MX15 Series, both hardware and software.
While not all or no options may be present on your specific unit, this section of the manual
incorporates the user documentation for all available options. There is no separate manual for
these options except possible manual addenda for special engineering request (SER) systems.
If your system has an SER number as part of the model number, refer to any manual addendum
that was shipped with the unit.

9.2 Option -HV: Additional AC Voltage Range

The -HV option provides an additional AC only output voltage range of 0-400 VRMS. There is
no equivalent 200 VRMS range associated with the -HV option but the standard 0-150 V RMS
and 0-300 V RMS remain available even if the -HV option is installed.
If the -HV option is installed, there will be three voltage ranges that can be selected from the
PROGAM menu. (150/300/400). Other than the range values, all other operations remain the
same. Note however that the -HV range is AC coupled and as such offers no DC output
capability. Thus, whenever the 400 V range is selected, the output mode is automatically set to
AC MODE and other modes cannot be selected.
Arbitrary waveforms
As stated before, the –HV (and –XV) range is AC (transformer) coupled and as such cannot
pass any DC component from the waveform generator. This means that any arbitrary waveform
created by the user that is to be used in the –HV range cannot have any DC offset. The MX
firmware will generate an error message if an attempt is made to select a custom waveform with
DC offset while in the –HV range.
MX15
133

Advertisement

Table of Contents
loading

Table of Contents