Olympus GX53 Instructions Manual page 57

Inverted system metallurgical microscope
Table of Contents

Advertisement

d
d
8
3
Place the socket part so that the heat radiating fin c faces down.
4
Loosen the burner clamping screws d (2 pcs.) of the socket part.
e
5
Grab the mercury burner e attached and remove the lower side from
the mount first, and then remove the upper side.
TIP
c
6
Hold the new mercury burner wrapped with gauze, etc. and attach
the + (positive) pole of the mercury burner e to the fixed mount on
the upper side, then attach the - (negative) pole to the mount on the
e
lower side.
NOTE
7
Tighten the burner clamping screws d (2 positions) of the socket part.
8
Align the external surfaces of the lamp housing with those on the
socket part, push the lamp housing straight downward and place the
socket part to the original position.
9
Tighten the clamping screw a using the Allen screwdriver.
a
10
Set the hour counter of the power supply for mercury burner (U-RFL-T)
to "0.0". For details, refer to the instruction manual provided with the unit.
USH-103OL: 300 hours
This value is based on the cycle of turning ON for 2 hours and
turning OFF for 30 minutes. Turning ON and OFF in the cycle
shorter than this will seriously shorten the service life of the burner.
The socket part is attached with either one of followings:
· Dummy burner for transportation (factory default)
· Old burner (to be replaced)
Be careful not to touch the burner directly with your hand and
avoid leaving fingerprints or stains on the burner. Otherwise,
the burner may be exploded due to the glass distortion
caused by stains. If fingerprints or stains are attached to
the burner, clean it by wiping gently with a piece of cleaning
paper slightly moistened with absolute alcohol.
Lamp service life
GX53
51