When the eyepiece is equipped with the eyepiece micrometer
1
While looking through the eyepiece equipped with the eyepiece
micrometer, rotate the diopter adjustment ring a to adjust so that the
scales or lines (A in the picture on the left) of the eyepiece micrometer
in the field of view are clearly visible. When you rotate the diopter
adjustment ring a , keep pressing the lower part b of the eyepiece.
2
Place the sample.
3
Engage the 10X objective in the light path. While looking into the
eyepiece equipped with the eyepiece micrometer, rotate the coarse
focusing knob and fine focusing knob to bring the sample into focus.
4
While looking through the eyepiece which is not equipped with the
eyepiece micrometer, rotate the diopter adjustment ring a to bring
the sample into focus.
When the observation tube is equipped with the interpupillary
distance adjustment ring c
Perform the same operation as described above.
Be sure to use the interpupillary distance adjustment ring c of the
observation tube instead of the diopter adjustment ring of the eyepiece
described above.