Attaching The Objective; Attaching The Revolving Nosepiece - Olympus GX53 Instructions Manual

Inverted system metallurgical microscope
Table of Contents

Advertisement

a
b
68
2
3
1

Attaching the objective

For attaching procedures, see "6-2 Replacing the objective" on page 47.

Attaching the revolving nosepiece

Be sure to screw in the cap to the (mounting) hole where
NOTE
the objective is not attached.
1
Loosen the revolving nosepiece clamping screw a using the Allen
wrench (opposite side: 3 mm). Then, align the revolving nosepiece
attached with objectives with the mount dovetail from the right side
and push it completely. If you attach the coded revolving nosepiece
at this time, pull out the cable b for the revolving nosepiece to the
left side of the microscope frame.
· Be careful, if the clamping screw is loosened too much, it
NOTE
may come off.
· When attaching or detaching the coded revolving
nosepiece, be sure to remove the cable from the
connector in advance. Also, if cables are fixed by the
cable holder, remove the cable from the cable holder in
advance.
2
While pushing the revolving nosepiece in the attaching direction with
your right hand, tighten the revolving nosepiece clamping screw a
to secure the revolving nosepiece.
3
If you connect the coded revolving nosepiece, connect the cable pulled
out to the left side of the microscope frame in
described below. For details, see "Connecting cables" on page 74.
· Relay unit for coded revolving nosepiece (GX-IFRES)
· Coded function system (U-CBS)
1
to either one of units

Advertisement

Table of Contents
loading

Table of Contents