Using The Mix Slider For Reflected Light Observation; Inserting The Mix Slider For Reflected Light Observation - Olympus GX53 Instructions Manual

Inverted system metallurgical microscope
Table of Contents

Advertisement

4-8 Using the MIX slider for reflected light observation

1
b
The following illustration shows the view without
stage in order to show the slider insertion slot
without obstacle.
e
38
1
a
2
c
d
f
Inserting the MIX slider for reflected light
observation
1 1
Loosen the mounting knob a at the rear right of the revolving
nosepiece and pull out the dummy slider b .
2
Insert the MIX slider for reflected light observation (U-MIXR) c into the
slider insertion slot of the revolving nosepiece so that the connector
d comes to the left side, and push it in the second level (position
where the clicking sound is heard).
MIX slider for reflected light
observation position
First level (pulled out)
Second level (pressed in)
3
Tighten the mounting knob a to secure the slider.
4
Connect the connector d to the MIX slider for reflected light
observation.
5
Set the main switch e of the control box (BX3M-CBFM) to (ON). The
buzzer sound is heard once.
In order to use the control box (BX3M-CBFM) with this
NOTE
microscope, the DIP switch must be set. For setting
procedures, see page 71.
You can check the state whether the MIX slider for reflected
TIP
light observation is engaged in or removed from the light path
by the POSITION indicator f of the hand switch (BX3M-HS).
Light path
OUT
IN