Meaning of notations in the following tables
Indicates the lighting state of the indicator of the control box (BX3M-CBFM).
: Turns ON.
: Blinks.
Problem
7. Control box (BX3M-CBFM) and hand switch (BX3M-HS)
a) Even though the objective is changed by rotating the revolving nosepiece, the OB indicator of the hand switch does not
turn ON.
For troubleshooting of the control box (BX3M-CBFM) not described in the above table, refer to the instruction manual for the
control box (BX3M-CBFM)
Repair request
If you cannot solve the problems even though taking actions described in Troubleshooting, please contact Olympus
for assistance. At that time, please tell them the following information as well.
· Product name and abbreviation (Example: Stage GX-SVR)
· Product number
· Problem
:Turns OFF.
: Varies depending on state.
Cause
The power of the control box
(BX3M-CBFM) is OFF.
The hand switch (BX3M-HS) is not
connected.
The hand switch (BX3M-HS) is
damaged.
The cable of the revolving
nosepiece or GX-IFRES is not
connected.
The objective is not correctly
engaged in the light path.
Remedy
Set the main switch to
(OFF) and
connect the control box to the AC
adapter of the control box again.
Then, set the main switch to (ON) .
Set the main switch to
 (OFF) and
connect the hand switch (BX3M-HS).
Contact Olympus.
Set the main switch to
(OFF) and
connect the cable of the revolving
nosepiece or GX-IFRES.
Set it to the click position properly.
GX53
Page
38, 76
38, 76
38, 68, 74
20
59