Olympus GX53 Instructions Manual page 55

Inverted system metallurgical microscope
Table of Contents

Advertisement

e
f
d
a
4
4
While pressing down the bulb fixing lever d , hold the halogen bulb
e wrapped with gauze, etc. and insert the terminal f until it touches
the pin position g .
Place the bulb fixing lever carefully to the original position to fix the bulb.
Be careful not to touch the bulb directly with your hand.
NOTE
If fingerprints or stains are attached to the bulb, wipe it
with a soft cloth, etc. to prevent the lifetime reduction or
g
the bulb explosion. If fingerprints or stains are attached to
the bulb, clean it by wiping gently with a piece of cleaning
paper slightly moistened with absolute alcohol.
5
Fit the halogen lamp housing from the above and while pressing down
the clamping screw a , tighten it.
3
Attaching the halogen lamp housing
1
Loosen the clamping screws a (2 positions) on the rear of the
b
microscope frame using the Allen screwdriver.
2
Insert the halogen lamp housing in the light source mount hole until
it touches the end.
Attach the lamp housing with the heat radiating fin b
CAUTION
facing up, and secure he sufficient space above, below
and at rear of the lamp housing. Otherwise, fire may be
caused.
3
Tighten the clamping screws using the Allen screwdriver.
Do not tighten screws too firmly. The lamp housing mount
NOTE
may be deformed.
4
Connect the cable for halogen lamp housing to the connector of the
power supply (TH4-100/TH4-200). For details, see "Connecting cables"
on page 74.
Applicable lamp
JCR12V100WB (made by Ushio Inc.)
Replacing the light guide halogen bulb for
1
transmitted light illumination
For procedures to replace the light guide halogen bulb
for transmitted light illumination, refer to the instruction
manual provided with the unit.
GX53
49

Advertisement

Table of Contents
loading

Table of Contents