Calibrating sensors
10.3
Inputs for calibration
10.4
Validation messages on calibration status
10.5
Offset calibration
10.6
Slope calibration
10.7
Deleting calibration
116
A keypad that enables you to input the real value you wish to use
for calibration.
In the special case of pH calibration with (a) buffer solution(s), the
value of the buffer solution must be entered. The same applies to
ORP.
This part is also used for the pH temperature compensation value,
if necessary, in the special case of the pH value.
The keypad is used in the same way as a normal numeric keypad.
If an incorrect value is entered or the value is outside the scale, the
value entered appears red and the "Save" button is disabled.
Confirmation message:
Before deletion, a confirmation message is displayed before
n
carrying out the process.
Valid calibration messages:
Calibration has been successfully completed.
n
Calibration has been successfully carried out.
n
Message of incorrect calibration:
Incorrect offset value. (Offset overflow for sensor permitted)
n
Incorrect slope value. (Permitted slope for the sensor has been
n
exceeded)
Two points too close. (2 points for pH calibration are not far
n
enough apart)
1.
Wait until the sensor has stabilised.
2.
Enter the calibration value.
3.
Press
'SAVE' .
1.
Wait until the sensor has stabilised.
2.
Enter the calibration value.
'SAVE' .
3.
Press
A confirmation prompt is displayed. There is no analysis of the
measured value; deletion can also be performed if the sensor is
unstable or missing.
'Delete' .
Press