Instructions for use KEY Laser III 1243
|
6 Safety checks

6.5 Spare unit leakage current test according to DIN VDE 0751-1

6.5 Spare unit leakage current test according to DIN VDE 0751-1
Requirement
Before the tester is started, the key switch of the KEY laser should be kept in the
on position until the test is complete.
▶ The spare unit leakage current test should be carried out between mains input
circuit and earth.
▶ The max. spare unit leakage current may not exceed 1.0 mA.
▶ Document the reading taken in the medical device record.
80/99