6 Safety checks
6.6 Further tests
6.6 Further tests
▪ Visual inspection of unit and accessories
▪ Function test
▪ No visible leaks in the unit
▪ Check internal safety systems
–
Emergency off switch
–
Key switch
–
Safety switch of swing arm (shutter lock)
–
Safety switch of input coupling of laser tube
–
Connection for external door contact
–
Connection for external warning lamp
–
Monitoring of cooling system
–
Energy monitoring/regulation
–
Shutter/absorber
–
Safety foot control
▪ Check external safety systems
–
Warning signs
Check whether the required warning signs are mounted at all entrance doors
to the KEY laser treatment room.
–
Warning lamps
Check that all entrance doors to the KEY Laser treatment room are equipped
with warning lamps.
–
Door switches
If door switches are installed, check that they are operating. If door(s) is/are
open, it should not be possible for the KEY Laser to switch to READY.
–
Safety goggles
Inspect all safety goggles to ensure they are fit.
81/99