Instructions for use KEY Laser III 1243
|
4 Operation
4.9 Modes
See also: 4.10 Threshold setting, Page 53

4.9.4 Feedback mode

The feedback mode combines the detection function of the detection mode with the
therapeutic function of the laser beam.
When the laser pulse trigger is pressed in the ready mode, laser pulses are triggered
only when, as a result of the detection, an altered tooth substance is found, i.e. when
the fluorescence measurement is above the said threshold value.
As soon as the fluorescence measurement falls below the threshold value, triggering
of laser pulses is stopped.
See also: 4.10 Threshold setting, Page 53