2 Setup methods according to DIN EN ISO 17664 | 2.1 References

2 Setup methods according to DIN EN ISO 17664

2.1 References

The reprocessing consists of the following basic steps:
▪ Cleaning and disinfection ( manually or automatically)
▪ Sterilisation
Note
Comply with national hygiene requirements, e.g. RKI guidelines.
Risk of injury during cleaning of the treatment unit.
Lack of instructions to the cleaning staff and lack of preparation of the treat-
ment unit can lead to the cleaning personnel sustaining injuries.
▶ Only trained professionals and instructed cleaning personnel may be
present in the treatment rooms.
▶ Position the chair for cleaning and turn the device off.
Product damage due to improper disinfection.
Malfunctions.
▶ Use the disinfectant according to the specifications of the manufacturer.
▶ No spray disinfection, perform wipe disinfection only.
▶ Do not immerse product or parts of the product in liquids.
▶ Mop up any spilled cleanser or disinfectant immediately.
Improper disinfection may lead to health hazards and material dam-
age.
Infection hazard to users and patients.
Damage to the sterile product.
▶ No hot air sterilisation, no chemical cold sterilisation, do not sterilise with
ethylene oxide.
Health hazard and property damage due to non-compliance with servi-
cing schedule.
Infection hazard to users and patients.
Product damage.
▶ Comply with servicing schedule.
Note
The preparation methods for the instruments and motors are found in the
separate instructions for use that accompany the instrument and motor
packages.
CAUTION
NOTICE
CAUTION
CAUTION
5 / 44
Care instructions Primus 1058 Life