4 Operation
4.17 Detector calibration
If the letter on the touchscreen does not correspond to the letter on the reference,
the corresponding reference should be obtained through the Customer Service
Center. For this purpose, it is essential to notify the contact person at KaVo of the
letter code of the reference value which is displayed on the touchscreen.
If the digit on the reference does not correspond to the digit on the touchscreen, the
digit on the touchscreen should be changed with the aid of the "Change reference
value" dialogue.
When the OK key is pressed, the reference value is accepted and you are returned
to the detector calibration menu.
The change is performed in the "Change reference value" menu with the aid of the
arrow keys
and
.
Note
The reference is located in a magnetically adhering plastic holder. It can thus be
fastened to the sheet metal cladding of the KEY Laser 1243.

4.17.2 Starting the calibration

Note
Have reference sample ready.
In the detector calibration menu, the calibration is started by pressing the START
key.
The user will be guided through the detector calibration by the program. You must
follow the instructions on the touchscreen.

4.17.3 Setting zero value

The following instruction appears on the touchscreen: "Direct handpiece at the floor"
Note
It should be ensured that the handpiece is not directed at a foreign light source.
67/99