Advertisement

3.4.7   Removing the tubehead back cover
Required tools: Hex key 3mm
1.
Remove the tubehead back cover by detaching the indicated screw and pull the cover outwards.
The screw has a screw cover cap over it.
2.
Remove the tubehead back cover.
3.4.8   Removing the rotating unit side covers
Required tools: Hex key 3mm
NOTICE! Turn the rotating unit carefully so that it does not collide with the lower shelf.
1.
Remove the screws (6 pcs) from the top of the rotating unit.
2.
Remove the screws (2 pcs) from below of the rotating unit.
3.
Pull the covers off.