Instructions for use KEY Laser III 1243
|
4 Operation

4.8 User test with hand piece 2060

4.8 User test with hand piece 2060
Note
Ensure that the laser tube is inserted into the input coupling and is screwed tight
and that the handpiece 2060 is mounted.
When the laser is in the ready mode the red pilot laser indicates the point of contact
of the therapy laser beam. This pilot laser makes it possible to judge whether the
transmission system is in the proper state.
See also: 4.6 Ready mode, Page 45
▶ Point the handpiece at a cotton wool ball or swab moistened with water.
If the pilot laser is visibly poor, broken up or completely invisible, the laser exit wind‐
ow is probably soiled or damaged.
See also: 5 Care, Page 72
▶ If the pilot laser is clearly detectable, then trigger a laser pulse using the trigger.
The typical noise which is heard when water is evaporated with the KEY Laser 1243
shows whether the laser exit is satisfactory.
Note
All functions and the laser energy of the laser system are constantly checked by
the control electronics. In the event of a malfunction, the automatic circuit breaker
of the KEY Laser 1243 is immediately activated.