6 Safety checks

6.2 Manual 'energy' safety check

6.2 Manual 'energy' safety check
The following test steps must be carried out on KEY Laser 1243 for this purpose:
▪ For each test step, you are required to calculate and document the mean value,
minimum, maximum, standard deviation and error (mean value to setpoint va‐
lue).
The error must not exceed +/- 10 %.
▪ The 'safety check report' page should be photocopied and completed in full. You
should then attach the form to the medical device record and customer service
report.
See also: 6.7 Safety check report, Page 82
Note
The only energy meters approved for making energy-related safety checks on the
KEY Laser 1243 are Gentec SUN EM1, Gentec PRJ-M RS-232 and OPHIR NOVA.
It must be ensured that these instruments are properly calibrated and inspected.
For suppliers, see servicing materials catalogue.
Refer to the technical information entitled 'Energy meter calibration for KEY Laser
1240/1242/1243'.