Instructions for use KEY Laser III 1243
|
9 Assembly notes

9.3 Electrical connection

9.3 Electrical connection
Note
The electrical protection and the supply cable must be implemented separately for
the KEY Laser 1243 in each case!

9.3.1 Local

▪ The connection must conform to the provisions of VDE 0100 for safety sockets.
▪ Receptacle connection for USA: 125V/30 AMP, NEMA 5-30R
▪ Fuse for 230V version with B16 or C16 automatic circuit breakers
▪ Fuse for 100-127V version with 25 A
▪ Fuse for 100-127V version with 20 A - Japan only
9.3.2 230V version at unit
▪ Mains cable, unit connector and mains connector requirements:
230V rated voltage
16 A rated current
Length: 3m
▪ Mains cable H05VV-F3 1.50 (3 x 1.5 mm²), not shielded
▪ Unit connector compliant with IEC 60320 C 19.
▪ Mains connector, even safety connector compliant with CEE (7) VII
▪ Protective conductor is important!
▪ All components used must comply with these requirements and the pertinent in-
country requirements.

9.3.3 100-127V version at unit

▪ Mains cable attached to an external voltage transformer or isolating transformer
▪ Mains connector, rated current with 25A (20A for Japan).
▪ Protective conductor is important!
▪ Mains connector compliant with specific in-country requirements
▪ Mains connector for USA only: 125V/30AMP, NEMA 5-30P, UL-, CSA-listed