Safe Laser Range; Warning Lamps - KaVo Key Laser III 1243 Instructions For Use Manual

Hide thumbs Also See for Key Laser III 1243:
Table of Contents

Advertisement

Instructions for use KEY Laser III 1243
|
2 Safety

2.4 Safe laser range

2.4 Safe laser range
Note
Each time the KEY Lasers 1243 is started, the regulations for operating lasers must
be observed.
The room must be identified by the required warning signs and lights at all points
of entry.
L
a
s
r e
B -
e
a
m

① Warning lamps

2.4.1 Warning lamps
In the planning stage, make sure that warning lights are installed at all entrance
doors to the treatment rooms in which the KEY Laser 1243 is operated.
If possible, install the electronics as follows:
The warning lights operate when power is applied to the socket for the KEY Laser
1243.
The warning lights must be installed by a professional electrician.
Note
The warning lamps are not included with the delivery of the KEY Lasers 1243.
② Warning signs