Instructions for use KEY Laser III 1243
|
2 Safety
2.5 Electromagnetic Transmissions
Immunity tests
Electrostatic discharge
(ESD) according to IEC
61000-4-2
Fast transient electrical
disturbances/ Bursts ac‐
cording to IEC 61000-4-4
Surges according to IEC
61000-4-5
Voltage interruptions,
short-term interruptions
and fluctuations of the
supply voltage according
to IEC 61000-4-11
Magnetic field with a sup‐
ply frequency (50/60 Hz)
according to IEC
61000-4-8
2.5.2 Guidelines and manufacturer's statement - Electromagnetic im‐
munity
The KEY Laser 1243 is for use in an environment like the one cited below. The
operator of the KEY Laser 1243 must ensure that it is operated in such an environ‐
ment.
IEC 60601 test level
± 6 kV contact discharge ±
8 kV air discharge
± 2 kV for power lines ± 1
kV for input and output li‐
nes
± 1 kV normal mode volta‐
ge ± 2 kV common-mode
voltage
< 5% V (> 95% drop in the
VT for ½ period 40%
V(60% drop in the VT for 5
periods 70% V(30% drop
in the VT for 25 periods<
5% V(> 95% drop in the VT
for 5 s (250 periods)
3 A/m
80 MHz – 2,5 GHz
NOTE: V T is the alternating mains voltage before the test level is used.
Conformance level
± 6 kV contact discharge ±
8 kV air discharge
± 2 kV for power lines ± 1
kV for input and output li‐
nes
± 1 kV push-pull voltage ±
2 kV common-mode volta‐
ge
< 5% V (> 95% drop in the
VT for ½ period 40%
V(60% drop in the VT for 5
periods 70% V(30% drop
in the VT for 25 periods<
5% V(> 95% drop in the VT
for 5 s (250 periods)
3 A/m
21/99
Electromagnetic environ‐
ment - guidelines
Floors should be made of
wood or concrete or have
ceramic tiles. When the
floor is made of synthetic
material, the relative humi‐
dity must be at least 30%.
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
The quality of the supply
voltage should correspond
to that of a typical business
or hospital environment.
When the user of the KEY
Laser 1243 needs conti‐
nued operation even when
the power supply is inter‐
rupted, it is recommended
to supply the KEY Laser
1243 from an uninterrup‐
ted power supply or a bat‐
tery.
Magnetic fields at the
mains frequency should
correspond to typical valu‐
es in a business and hos‐
pital environment.