Advertisement

Section 4
Service

Spare Fuses

4-4
MaxQ SHKE8000 Series Shakers
Two fuses are located on the front of the relay enclosure located in the
lower right area of the stackable cabinet. Refer to Table 4-2 for a list of
thier electrical ratings, part numbers and application. Gain access to the
relay enclosure by removing the kick panel on the lower front of the
cabinet. It is held in place by a screw on each end of the panel and two
tabs at the top. To access the fuses, remove the retainer/cover plate holding
the relay enclosure in place. See Figure 4-6.
Table 4-2. Fuses
MaxQ SHKE8000 Series Shakers
Rating
Application
0.25 amp
Main Power Relay Board
1.6 amp
Drive Motor
Caution Do not substitute! Replace these fuses only with fuses of identical
s
electrical ratings.
Figure 4-6. Relay Enclosure with and without Cover Plate
Part Number
230144
230145
Thermo Scientific