Three temperature alarms are programmed into the MaxQ SHKE8000
Series Stackable Orbital Shaker; Overtemperature, Undertemperature
(Model SHKE8000-7 & SHKE8000-8CE), and Tracking Limits.
Set the Overtemperature Alarm
The Overtemperature alarm activates whenever the operating temperature
goes above the Overtemp setpoint temperature by a few tenths of °C. This
adjustable limit is set at the factory at approximately 64°C. To change this
value, open the Configuration menu by
pressing the down arrow, the up arrow, and
the Silence button, and then the down arrow
4
3
2
1
once again, in the sequence shown at right.
Press
Then press the Temperature
button beneath Alarms.
To change the Overtemperature
alarm setting, press the Time
button beneath Over.
The following screen appears and
the present Overtemperature
alarm setting flashes.
Change the Temperature setting by pressing the up or down arrow. When
set, press any of the three buttons (Time, Speed, or Temp) to save the new
setpoint value and return to the previous screen or press nothing for about
15 seconds to save the new setpoint value and return the display to the
Operating Screen.
Note The Overtemp Alarm setpoint values are calculated from the
hardware and will not include every numerical value between the upper
and lower limits.