Handle Adjustment - Thermo Scientific MaxQ SHKE8000 Series Operating And Maintenance Manual

Stackable orbital shakers
Hide thumbs Also See for MaxQ SHKE8000 Series:
Table of Contents

Advertisement

Handle Adjustment

Thermo Scientific
1. Grasp the platform handle and pull up on the inner release bar to
move the handle to a horizontal (down) position.
2. Remove the screws from the sides of
the handle, as indicated in Figure 4-
12. Remove the handle.
3. Turn the adjustment screws just one-
half turn clockwise to increase the
clamping force, or one-half turn
counterclockwise to reduce the
clamping force.
4. Reassemble the handle and install the side screws. Slide the platform
into the shaker chamber. Place the handle into a vertical position.
5. Close the door and with the unit empty, set the speed to 300 RPM. If
there is undue vibration noise, repeat the steps above. If not, return the
unit to service.
Figure 4-12. Adjustment
MaxQ SHKE8000 Series Shakers
Section 4
Service
4-7