Remote Alarm System

Cycle Complete

Power Failure

2-16
MaxQ SHKE8000 Series Shakers
Most of the alarm states described previously (Table 2-2, pg 2-7) can alert
a remote alarm monitoring system through an internal SPDT relay
connected to an RJ-11 connector on the left front of the shaker, behind
the kick panel. Refer to 'Connect the Remote Alarm' in Section 1. For the
convenience of the laboratory, these remote alarms can be individually
turned on or off. Any of the remote alarms that are set to On will activate
the internal relay.
Note The Overtemp Shutdown, Undertemp Shutdown, Platform Stalled,
and Check Fuse remote alarms cannot be deactivated.
To set the remote alarms to On or Off, open the Configuration menu by
pressing the down arrow, up arrow, and the Silence button, and then the
down arrow three times, in the sequence shown below.
The screen shown at the right
will appear on the display.
Press the Speed button beneath
Remote (Rmte). The alarms will be shown in the following sequence:
Toggle the Cycle Complete alarm
with either the up (On) arrow or
the down (Off) arrow. Pressing
the Temp button beneath Next
advances the display to the next alarm, saving the Cycle Complete alarm
setting to memory. If no buttons are pressed, the display will automatically
return to the Operating Screen after about 15 seconds, saving the selection
to memory.
Toggle the Power Failure alarm
with either the up (On) arrow or
the down (Off) arrow. Pressing
the Temp button beneath Next
advances the display to the next alarm, saving the Power Failure setting to
memory. If no buttons are pressed, the display will automatically return to
the Operating Screen after about 15 seconds, saving the selection.
Actual
Cycle Complete:
Setpoints
Actual
Power Failure:
On