Section 2
Operation
Set the Temperature Alarm
The Temperature Tracking alarm activates whenever the operating
temperature goes above or below the setpoint temperature by a user
Tracking Limit
selectable value in the range of 1°C to 20°C. The limit is set at the factory
as 10°C above and below the temperature setpoint.
Note The above and below limits are always the same value.
To change this limit, open the
Configuration menu as in the previous
alarm procedures, by pressing the down
4
3
1
2
arrow, up arrow, and the Silence button,
Press
and then the down arrow once again, in the
sequence shown at right.
Then press the Temperature
button beneath Alarms.
To change the Temperature
Tracking alarm setting, press the
temperature button beneath
Tracking (Trcking).
The screen at right appears and
the present Temperature
Tracking alarm limit setting
flashes.
Change the Temperature Tracking limit by pressing the up or down
arrow. When set, press any of the three buttons (Time, Speed, or Temp)
to save the new setpoint value and return to the previous screen, or press
nothing for about 15 seconds to save the new setpoint value and return the
display to the Operating Screen.
Thermo Scientific
MaxQ SHKE8000 Series Shakers
2-15