Calibrate Platform Speed (Rpm) - Thermo Scientific MaxQ SHKE8000 Series Operating And Maintenance Manual

Stackable orbital shakers
Hide thumbs Also See for MaxQ SHKE8000 Series:
Table of Contents

Advertisement

Section 4
Service
Calibrate Platform
An external calibrated speed measuring device can be used to adjust the
actual platform speed so that the unit setpoint speed matches the external
Speed (RPM)
device measurement. Calibration of the platform speed is performed at the
factory at 250 RPM. To change the actual platform speed, open the
Configuration menu by pressing the down arrow, up arrow, and the
Silence button, and then the down arrow once again, in the sequence
shown below.
Then press the Speed button
beneath RPM.
The value shown on this screen is
the present Speed setpoint. Using
the up and down arrows, increase
or decrease the platform speed
until the reading on an
independent, accurate speed measuring device matches the shaker speed
setpoint.
When finished, press any of the three buttons (Time, Speed, or Temp) to
save the new setpoint value and return to the previous screen, or press
nothing for about 15 seconds to save the new setpoint value and return to
the Operating Screen.
Thermo Scientific
Model SHKE8000 Series Shakers
4-9