Single Unit On A Stand; Positioning - Thermo Scientific MaxQ SHKE8000 Series Operating And Maintenance Manual

Stackable orbital shakers
Hide thumbs Also See for MaxQ SHKE8000 Series:
Table of Contents

Advertisement

Single Unit on a

Positioning

Thermo Scientific
Stackable Shaker Stands (Stock #194222 and 194223) are shipped with
the following materials:
Stand
4 - Foot pads
4 - Painted spacers blocks 1" high
4 - ½"-13 x 4" hex head bolts
4 - 5/16" lockwashers
4 - D-shaped washers
Locate the shaker stand on a
firm, level surface in an area
free of dust and dirt. To allow
for service access, the shaker
back must be at least 4 inches
from the wall. Note location
of power and light switches
(Figure 1-9).
As the electrical plug is the
mains disconnect for the
unit, electrical wall outlet
must be easily accessible at all times.
Warning Use extreme caution when lifting and positioning these units.
Model SHKE8000 & SHKE8000-1CE each weigh 550 lbs (249.5kg) and
Model SHKE8000-7 & SHKE8000-8CE each weigh 600 lbs (272.2kg).
Warning Do not lift the unit by hand. Always use suitable equipment
designed to support over 600 lbs. (272.2kg).
Warning It is imperative that all specified installation hardware be used
when stacking shakers. Failure to do so may result in severe injury and/or
equipment damage.
Caution Do not lift Model SHKE8000-7 or SHKE8000-8CE under the
refrigeration assembly attached to the side of the shaker! This structure is
not designed to support the weight of the unit.
Caution Do not lift or position the unit by the door or door handle.
Damage to door mounting and gasket may occur.
Actual
Setpoints
Silence
Time
Speed
Temp °C
Figure 1-9. Model SHKE8000 Series Incubated
Stackable Orbital Shaker
s
START
(arrow
keys)
STOP
rbital Shaker
Light
Switch
s
s
s
MaxQ SHKE8000 Series Shakers
Section 1
Installation
Power
Switch
s
1-5