Install the shaker on a sturdy table or bench. The support structure must
be able to accommodate both the weight and operating motion of the
shaker. The shaker can also be placed under a bench or in a "knee well."
Note If your platform is already installed, skip this section and continue
to next section.
Caution Remove the shipping bracket and install the shaker platform
before plugging in or attempting to operate the unit.
After removing the orbital mechanism
shipping bracket and installing the
platform, remove this protective decal from
the control panel to begin shaker operation.
Figure 1-3. Orbital Mechanism Shipping Bracket
To protect the shaker's orbital mechanism during shipment, a sheet metal
shipping bracket is installed and must be removed before the unit can be
Bracket
operated. Using a 7/16" and 9/16" hex wrench, remove the three 1/4"
and the single 3/8" screws. Retain this hardware for future shipping.