Th Pa Np Pu Am Cm Bk Cf Es Fm Md No Lr Thorium Protactinium Uranium Neptunium Plutonium Americium Curium Berkelium Californium Einsteinium Fermium Mendelevium Nobelium Lawrencium Thermo Scientific SSRZ Sample Holder Instruction Notice AA83858 - A • June 2019...
General Safety Guidelines ................... Hazard Statements and Symbols ................. The Sample Holder Description ............... Installation ....................... Install SSRZ Sample Holder in the ARL EQUINOX ..........Connect the Sample Holder to the Power Supply ..........Set the Incidence Angle ..................Sample Preparation Procedure..............
The information in this document is subject to change without notice. We assume no responsibility for any errors that may appear in this document. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 3...
If you need any help, please contact the local Thermo Fisher Scientific representative. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 4...
Page 5
Chapter | Introduction Manufacturer THERMO FISHER SCIENTIFIC INEL SAS 71 rue d’Orléans 45410 ARTENAY, FRANCE Tel : +33(0)2-38-80-45-45 Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 5...
Transport of the sample holder should be done with extreme care. • Never operate a damaged sample holder. • Never operate the sample holder with a damaged power cord. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 6...
It may also be used to alert against unsafe practices. SERVICE! A Service message indicates that the described procedure should be performed by qualified personnel only. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 7...
4 - Level indicator. 5 - Sample cup for analysis. 6 - Height adjustment screw. 7 - Power cord (12V). 8 - Platform for bulk sample Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 8...
Page 9
8 - Platform for bulk sample SERVICE! Do not touch the clamping screws (3). This operation has to be done by service personnel only. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 9...
Chapter | Installation Installation 5.1 Install SSRZ Sample Holder in the ARL EQUINOX Insert the sample holder in the sample holder mounting hole of the instrument, do not use any unnecessary force. The sample holder and the mounting hole (Ø 30 mm) are designed to avoid any mechanical clearance resulting in irreproducible results.
2 - Incidence angle equal to 0° 3 - Medium line 10° of incidence angle 4 - Short line 2° of incidence angle Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 11...
The surface of the powder must not exceed the level of the surface of the powder mount, in order to avoid line displacement error due to the height error. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 12...
Chapter | Sample Preparation Procedure 6.2 Insert the Sample Cup in the SSRZ Use the height adjusting screw (6) to set the right alignment for the beam. Use a glass slide to protect your sample as shown on pictures beside.
Page 14
Chapter | Sample Preparation Procedure When the level is reached, rotate the Level indicator. Remove slowly the glass slide. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 14...
To change the sample height, unscrew the two screws (1), move the sample support to the right level and secure it using the screws. Set up in the highest position. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 15...
Page 16
When the level of the sample is reached, rotate the level indicator. Remove slowly the glass slide. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 16...
Chapter | Maintenance and Cleaning Maintenance and Cleaning Sample holder can be cleaned with a soft and dry brush. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 17...
Check your sample preparation, make sure that you have right amount of powder in the cup. • Check the alignment of the sample holder. Confidential, limited rights Instruction Notice, Sample Holder SSRZ PN AA83858 | Revision A | 13-Jun-2019 Page 18...
Chapter | Appendix 1: Conformity Certificate Appendix 1: Conformity Certificate Thermo Fisher Scientific has certified by internal assessment that the SSRZ sample holders comply with the CE mark according to the following standards and directives: • EN 61010-1 (2010): Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements, •...
Connect the Sample Holder to the Power Supply • General Safety Guidelines • 6 Hazard Statements and Symbols • 7 Insert the Sample Cup in the SSRZ • 13 Install SSRZ Sample Holder in the ARL EQUINOX • 10 Installation • 10 Introduction •...